26 Agilent X-Series Signal Generators User’s Guide
Signal Generator Overview
Rear Panel Overview (N5173B & N5183B)
15. Z AXIS OUTPUT
This female BNC connector supplies a +5 V (nominal) level during retrace and band- switch intervals
of a step or list sweep. During step or list sweep, this female BNC connector supplies a - 5 V
(nominal) level when the RF frequency is at a marker frequency and intensity marker mode is on.
This signal is derived from an operational amplifier output so the load impedance should be greater
than or equal to 5 kohms.
Connector female BNC Impedance nominally 50 Ω
Signal A nominal signal level greater than 4 dBm.