Chapter 4 CATALYS® System Procedure
Page 190 of 314 Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Locking the Suction Ring/Disposable Lens Assembly to the Laser
After the system verifies suction ring capture, the Lock panel opens, and you are prompted to
lock the suction ring/disposable lens assembly to the laser, as follows:
1. Using the joystick control on the Mobile Patient Bed or patient chair, adjust the bed/
chair so that the suction ring/disposable lens assembly is in the Lock Zone and the lateral
and vertical forces are within the green zones, as shown in the following figure.
QUICK TIP: LOCKING THE SUCTION RING
Monitor the Lateral Force Sensor, and make proper adjustments with the joystick to
minimize lateral forces.
Figure 4.34 Adjust Patient Positioning
Lock Zone
Lateral and Vertical
Forces Must Be
Within Green Zones
;
Move Bed/Chair
Opposite to
Indicated Forces