Appendix A – Chapter 1 Mobile Patient Bed System Description
Page 246 of 314 Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
1.3 Mobile Patient Bed Specifications
* OPTIMEDICA reserves the right to make changes to the device(s) herein. Device(s), therefore, may not
agree in detail with the published design or specifications. All specifications are subject to change with‐
out notice. Contact OPTIMEDICA or your local OPTIMEDICA representative for information on changes
and new products.
Electrical Requirements
Voltage 100 – 240 VAC
Frequency 50/60 Hz single-phase
Current 270 VA
Environmental Requirements (Operating)
Maximum altitude 7,000 ft (2,134 m)
Temperature range 59 – 89.6 °F (15 – 32 °C)
Maximum humidity Up to 80% @ 89.6 °F (32 °C) non-condensing
Operating environment
Temperature controlled environment with temperature controlled to
within ±3.6 °F (±2 °C)
Atmospheric pressure No atmospheric pressure dependency during the operation of the device.
Environmental Requirements (Non-operating)
Maximum altitude Standard commercial shipping altitude
Temperature range 14 – 131°F (-10 – 55°C)
Maximum humidity Up to 80% @ 131°F (55°C) non-condensing
Physical Characteristics
Length (bed in flat position and
headrest frame fully extended)
78 in ±2 in (1.98 m ±.05 m)
Width 30 in ±1 in (.76 m ±.025 m)
Weight 340 lb ±20 lb (154 kg ±9 kg)
Maximum patient weight 399 lb (181 kg)
Safe working load 439 lb (199 kg)
Total mass 778 lb (353 kg)
Articulation Range
Backrest inclination 0° to 85°
Seat inclination 0° to 20°
Footrest inclination 0° to 80°
Bed height (from floor to top of seat
cushion)
Lowest position: 20 in (51 cm)
Highest position: 34 in ±2 in (86 cm ±5 cm)
Classifications
Floating Class II
IEC 60601-1 protection against elec-
tric shock
Type BF
IEC 60601-1 patient leakage Footrest, seat, backrest, headrest, armrests