Appendix A – Chapter 4 Mobile Patient Bed Safety and Regulatory
Page 280 of 314 Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
No modification of this equipment is allowed.
Use of high frequency surgical equipment, cardiac defibrillators, and defibrillators‐monitors may
cause unintended motion of the bed. Do not use such equipment while performing ophthalmic
procedures.
Improper use of high frequency surgical equipment may lead to unintended patient burns. Refer
to the high frequency surgical equipment manufacturer’s operator manual for safe operation
and compatibility with the Mobile Patient Bed.
Use of the Mobile Patient Bed with other equipment may lead to increased leakage current,
resulting in electrical shock. Refer to the equipment manufacturer’s operator manual for safe
operation and compatibility with the Mobile Patient Bed.
CAUTIONS
Federal law (USA) restricts this device to sale by or on order of a physician.
Read this operator manual thoroughly and be familiar with its contents prior to using this
equipment.
When the Mobile Patient Bed is interconnected with other medical electrical equipment, leakage
currents may be additive. Ensure that all systems are installed according to the requirements of
IEC 60601‐1.