Cell holder
Standard cell holder, 50 mm
SPECORD S600 accessories Edition 09/2014 11
Inserting the cell correctly
1. Fill the cell with your sample at least 20 mm high.
2. Insert the cell tight against the contact surface of the cell holder (arrow in
Fig. 2-3).
Fig. 2-3 Cell inserted correctly in the cell holder
Note
One of the decisive factors for extinction measurement is the path length. It is
therefore important to place all samples during a measurement into the same
position and orientation. This prevents wedge or angle errors or the parallel
offset of the radiation, causing measuring errors.