Cell holder
Holder for 100 mm absorption tube
SPECORD S600 accessories Edition 09/2014 23
10 Holder for 100 mm absorption tube
Absorption tubes can be used to measure liquid or gaseous samples for which
a path length of 100 mm is required. The tube holders position the absorption
tube to the beam path. In addition, two universal holders are required.
Layout
1 Holder
2, 7 Quartz window
3, 6 Mounts with sealing rings
4 Plugs
5 Glass body of the absorption tube
Fig. 10-1 Absorption tube with holders
Inserting holders and absorption tube into SPECORD S600
1. Place the first universal holder onto the support rods. The adapter plate
points to the lamp housing.
2. Place the second universal holder onto the support rods the adapter plate
pointing to the polychromator.
3. Slide one holder each from the top onto the adapter plates of the universal
holders.
4. Move the second universal holder to allow the absorption tube to rest on the
holders.
5. Place the absorption tube onto the holders.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Analytik Jena SPECORD S600 and is the answer not in the manual?

Analytik Jena SPECORD S600 Specifications

General IconGeneral
BrandAnalytik Jena
ModelSPECORD S600
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals