Peltier-tempered accessories
Peltier tempered 6- and 8-cell changers
SPECORD S600 accessories Edition 09/2014 59
Window Cyclic temperature mode
Fig. 15-17 Measuring parameter configuration for Peltier-tempered cell holders -
temperature control: Cyclic
The table in the window Cyclic temperature mode contains the temperatures
to be activated, the waiting time between reaching the temperature and the
start of the measurement, and the option to start a measurement at the
respective temperature stage (Opt. measurement). The number of activated
temperature values is entered in the input field Count of temp. values. The
table rows are created in accordance with this number.
The parameters of the temperature stages can be edited individually in the table
rows or defined automatically using the checkboxes and input fields.
Parameter /
button
Description
Temperature
accuracy
Set the accuracy of the temperature control.
Range: 0.1°C to 2°C
Count of temp.
values
Number of temperatures to be activated.
Constant waiting
period
If enabled, the same waiting period will apply between
reaching the target temperature and the measurement
for each temperature stage. The waiting period must be
entered in the adjacent input field.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Analytik Jena SPECORD S600 and is the answer not in the manual?

Analytik Jena SPECORD S600 Specifications

General IconGeneral
BrandAnalytik Jena
ModelSPECORD S600
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals