Peltier-tempered accessories
Temperature control unit for Peltier-tempered accessories
SPECORD S600 accessories Edition 09/2014 63
Fig. 15-21 Temperature curve in the block and inside the cell for different cells
Recording the inside cell temperature
1. Close a cell with the opening for round plugs with the cell sensor supplied
and slide the connection socket onto the cell sensor.
2. Turn the change-over switch at the control unit to "cell".
3. If the cell sensor is no longer required pull the cell sensor out of the
connection socket.
Error message!
If the temperature display in the control unit flashes, either the cell sensor is
not properly connected to the connection socket or the sensor is faulty.
Changing fuses
If the fuses in the temperature control unit are faulty you can replace them.
1. Disconnect the mains cable from the control unit connection.
2. Open the fuse holder (5 in Fig. 15-19
3. Replace the faulty mains fuses. Use the following fuses:
2 x T 2.5 AH / 250V, type 19181 by Wickmann
4. Close the fuse holder again.
5. Plug the mains cable into the connection of the control unit and the mains
socket.
6. Switch on the SPECORD S600.