EasyManuals Logo

Analytik Jena SPECORD S600 User Manual

Analytik Jena SPECORD S600
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
Peltier-tempered accessories
Peltier tempered 6- and 8-cell changers
58 Edition 09/2014 SPECORD S600 accessories
To activate and hold a constant temperature, select the mode Fix and carry out
the following configuration.
Parameter /
button
Description
Temperature
accuracy
Set the accuracy of the temperature control.
Range: 0.1°C ... 2°C
Fix
temperature
Set target temperature.
Waiting time
before Measure
Waiting period after reaching the target temperature until
the start of the measurement.
Cyclical temperature mode
In this temperature mode you can activate up to 200 temperature values. The
measurement is taken each time when the temperature stage has been
reached. Select the mode Cyclic in the window Peltier temperature-control-
led. Click on the Setup button and carry out the following configuration in the
window Cyclic temperature mode:
Window Peltier temperature-controlled
Fig. 15-16 Selection of Temperature control: Cyclic