Peltier-tempered accessories
Peltier temperature-controlled cell holder, air-cooled
SPECORD S600 accessories Edition 09/2014 43
To activate and hold a constant temperature, select the mode Fix and carry out
the following configuration.
Parameter /
button
Description
Temperature
accuracy
Set the accuracy of the temperature control.
Range: 0.1°C ... 2°C
Fix
temperature
Set target temperature.
Waiting time
before Measure
Waiting period after reaching the target temperature until
the start of the measurement.
Simultaneous temperature control
Fig. 15-6 Measuring parameter configuration for Peltier-tempered cell holders -
temperature control: Simultaneous
For measurement recording during the heating or cooling from the start to the
end temperature with a preset heating or cooling rate set the mode to
Simultaneous:
Parameter / button Description
Start temperature
Temperature during the 1st measurement
End temperature
Temperature during the last measurement
Measure every... °C
Temperature interval during which a measurement
should take place.
Gradient
Heating or cooling rate
Cuvette
measurement
sensor active
If enabled the temperature control is via the cell
sensor directly in the sample, otherwise via the sensor
in the cell block.
Cooling down after
ramp
If enabled the sample is moved to and held at this
temperature after the end temperature has been
reached. The standby temperature must be entered in
the adjacent input field.