Peltier-tempered accessories
Peltier tempered 6- and 8-cell changers
SPECORD S600 accessories Edition 09/2014 55
4. If the subsequent measurement is carried out with standard cells or semi-
microcells carry out an adjustment with an empty cell changer.
If microcells are used during the subsequent measurements, place a water-
filled cell in each of the 6 positions for the adjustment.
5. Start the automatic adjustment by clicking on the button [Adjustment].
Using the cell sensor
The temperature control is optionally carried out via the sensor in the cell block
or the cell sensor.
Caution! Fragile sensor!
Do not use force when placing the sensor. It is not necessary to press the
sensor in place because it receives enough pressure from the contact pins in
the cover cap.
Caution! Flashing temperature display!
If the temperature display in the control unit (Fig. 15-20 p. 62) flashes, either
the cell sensor is not properly connected to the connection socket or the
sensor is faulty.
1 Cell sensor
2
Cover cap of the cell holder
3
Set screw
Fig. 15-12 Cell sensor for Peltier-tempered cell changers
1. Seal a standard cell with round plug with the cell sensor supplied.
2. Place the cell into the cell block and close it with the cover cap.
3. Turn the change-over switch at the control unit to "cell".
Removing the cell changer from the sample chamber
1. Switch off the heat exchanger and temperature control unit and disconnect
them from the mains supply.
2. Move the cell changer into position 5. The cell block is centered on the base
plate.
3. Disconnect the connector from the connection on the rear of SPECORD
S600.
4. Pull up the cell changer on the rear side of the base plate until it disengages
from the clamp with a click. Lift the cell holder out of the sample chamber.
5. Disconnect the hoses from the heat exchanger. Collect the draining liquid
6. Pack the cell changer in accordance with Fig. 15-13.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Analytik Jena SPECORD S600 and is the answer not in the manual?

Analytik Jena SPECORD S600 Specifications

General IconGeneral
BrandAnalytik Jena
ModelSPECORD S600
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals