Peltier-tempered accessories
Heat exchanger for Peltier-tempered accessories
66 Edition 09/2014 SPECORD S600 accessories
Note
When placing the heat exchanger note that the maximum delivery height is
approx. 1.2 m.
Dependent on the location of the heat exchanger, the routing of the hoses
and for design reasons air cushions may form during the system filling
process and only dissipate after a few minutes. Brief switching off and on can
accelerate the dissipation of this air cushion.
Replacing the fuses at the heat exchanger
If the fuses in the heat exchanger are faulty you can replace them.
1. Disconnect the mains cable from the heat exchanger connection.
2. Open the fuse holder by pulling its cover.
3. Replace the faulty mains fuses. Use the fuses in accordance with the
details in the "Technical data".
4. Close the fuse holder again.
5. Plug the mains cable back into the connection of the temperature control
unit and the mains socket.