5-4 ISDB-T and BER Verification, Options 30 and 79 Option Verification
5-14 PN: 10580-00255 Rev. J MT8212E and MT8213E MM
5. Set the mode of the MT821xE to ISDB-T Signal Analyzer and preset the instrument.
6. Press the Frequency/Level main menu key and confirm that the instrument is set to Channel 13. Change
the Reference Level to –10 dBm and ensure that the Pre Amp is Off.
7. Press the Meas Selection main menu key and press Phase Noise (red dot appears on label).
8. Press the Meas Setup main menu key and then the Meas Mode submenu key. Use the Down Arrow key
to select Average and press the Enter key.
9. Wait until Average counter displays (10/10).
10. Record the 10 kHz and the 100 kHz phase noise readouts in Table A-29, “ISDB-T Signal Analyzer Phase
Noise” on page A-22.
11. Record the Frequency Error in Table A-29.
12. Set the frequency of the MG3700A to 623.14285714 MHz and change the MT821xE Channel to 38.
13. Wait until the Average counter displays (10/10).
14. Record the 10 kHz and the 100 kHz phase noise readouts in Table A-29.
15. Record the Frequency Error in Table A-29.
16. Set the frequency of the MG3700A to 767.14285714 MHz and change the MT821xE Channel to 62.
17. Wait until Average counter displays (10/10).
18. Record the 10 kHz and the 100 kHz phase noise readouts in Table A-29.
19. Record the Frequency Error in Table A-29.