5-16 DVB-T/H Signal Analyzer Verification, Options 64 and 57 Option Verification
5-80 PN: 10580-00255 Rev. J MT8212E and MT8213E MM
24. Use the rotary knob to highlight Moving Average and press the Enter key.
25. Press the Average Count soft key, then enter 10, and press the Enter key.
26. Wait until Average (10/10) appears at the top of the display.
27. Record the Frequency Error in the “474.09 MHz” row of Table A-72, “Option 64, DVB-T/H Signal
Analyzer, Frequency Lock Range” on page A-46.
28. On the MG3700A, set the frequency to 473.91 MHz.
29. Press the Execute Measure main menu key. After Average (10/10) appears, record the Frequency Offset
in the “473.91 MHz” row of Table A-72, “Option 64, DVB-T/H Signal Analyzer, Frequency Lock Range”
on page A-46.
Level Accuracy Verification
The tests in this section verify the level accuracy of the Cell Master in DVB-T/H Signal Analyzer mode.
Equipment Required:
Anritsu MG3700A Vector Signal Generator
Power Meter, Anritsu Model ML2438A
Programmable Attenuator, Anritsu Model MN63A
RF Power Amplifier, Mini Circuits Model TIA-1000-1R8
Power Sensor, Anritsu Model MA2482D (2)
Power Splitter, AeroflexWeinschel Model 1870A
Fixed Attenuator, Aeroflex/Weinschel Model 44-10
RF Coaxial Cable, Anritsu Model 15NN50-1.5C (3)
Adapter, Anritsu Model 34NN50A
Midwest Microwave ADT-2615-NF-BNM-02 Adapters (2)
Anritsu 1N50C RF Limiter
Anritsu PN 2000-1627-R Coaxial Cable