5-18 Mobile WiMAX Signal Analyzer Verification, Options 66 and 67 Option Verification
5-96 PN: 10580-00255 Rev. J MT8212E and MT8213E MM
e. Using the Down Arrow key, step through the selection list until mWiMax is highlighted.
f. Press the Set key.
g. Press the F6 (Return) soft key.
h. Press the Set key. The Select Package list box appears. Again select mWiMax and press Set.
i. Another file list appears. Select (highlight) 5m512g8_2_5_cap.
j. Press the Set key.
k. Press the MOD On/Off key and verify that the LED is On. Confirm that the “playing” indicator is
displaying the moving pattern.
l. Press the Frequency key, then enter 2600.5 MHz.
m. Press the Level key, then enter 2dBm. Turn the output On.
22. Set the calibration factor frequency of the power sensor to 2600.5 MHz.
23. Adjust the MG3700A level setting with the knob so that the power meter reads –15.0 dBm ± 0.2 dB.
24. Set the MT821xE to Mobile WiMax Signal Analyzer mode and preset the instrument.
25. Set the MT821xE as follows:
a. Press the Freq main menu key and set the Center Frequency to 2600.5 MHz.
b. Press the Setup main menu key and set the Bandwidth to 5MHz.
c. Press the Frame Length submenu key and set the Frame Length to 5ms.
d. Press the Measurements main menu key and press RF Measurements, then press Power vs Time.
26. Repeat Step 9 through
Step 20, recording the results into the test record in Table A-86, “Option 66,
Mobile WiMAX Channel Power Accuracy (5 MHz Bandwidth and 5 ms Frame Length)”.