102 ATOMLAB 500 DOSE CALIBRATOR
Figure 5.24. The Moly Assay reading screen with Moly Background Reading Activity displayed.
11. Select <Enter> to record the Moly Background.
12. Place the vial of Tc-99m into the Moly Shield and put it into the detector.
13. Select <Record Reading>. The system will prompt you to wait 30 seconds the activity reading.
NOTE: You can select <Record Reading> to recount the Moly if desired.
14. Select <Enter> to return to the Moly Assay screen to count the vial or syringe of Tc-99m in the
Moly shield.
15. Remove the Tc-99m vial or syringe from the Moly Shield and place the vial or syringe into the detector.
16. Select <Record Reading> to record the Tc-99m activity. The Moly Assay results screen is now displayed
giving the Moly concentration of µCi Mo/mCi Tc-99m.
17. Touch <Enter> to return to the Enter Kit # Data screen.
NOTE: Touch <Print Label> if a label is desired.
18. To save the data, touch <Save Data> and then <OK> to confirm. If you have printed a label
already, the system returns to the Nuclear Pharmacy screen.
19. If you have not yet printed a label, a screen is displayed at this point allowing you to print a
label or skip printing. Touch <Print Label> if a label is desired or <Skip Printing> if you do
not want to print a label. After printing or skipping the label, the system returns to the
Nuclear Pharmacy screen.