6. Repeat step five to record a total of three readings. The system calculates the average
reading and compares it to the calculated standard, then determines the average variance.
NOTE: If any of the readings are out of the acceptable range they will be highlighted in red.
7. Touch <Save Data>. The save Accuracy Test data screen is displayed.
8. Touch <Staff Member>. The Staff Member List is displayed. Touch to highlight any listed
staff member. (If only one person is on the list, the system will default to that
person on the Save Geometry Test Data screen.) Touch <Enter> to select the
staff member and return to the Save Geometry Test Data screen. If necessary, touch
<Add Staff Member> or <Delete Staff Member> to update the Staff Member List.
9. Touch <Measure> to go to the Measure screen. A message is displayed: "if you proceed,
recorded data will be lost, is it OK to exit and abort?". Touch <Yes> to delete the test and go
to the Measure screen, or <No> to return to the study.
10. Touch <Back> to return to the Accuracy Test screen without saving.
11. Touch <Save Data> to save the test just performed and advance to the save Accuracy Test
data screen.
12. Touch <View Data> to view this test and previously performed Accuracy Tests.
13. Touch <Back> twice to return to the QA Dose Calibrator screen.
14. If you select <View Data> you can see the Accuracy Test just performed, or any Accuracy
Test previously completed. If a printout is desired, touch <Print> to print the Accuracy Test
data screen.