Biodex Medical Systems, Inc. © 2016 167
Figure 6.37. The Semi-Automated Decay Test Ended screen.
NOTE: When in the manual portion the system works like the Manual Decay Test.
14. Touch <End Test> to end the test and advance to the normalized reading screen.
15. Touch <Select Normalized Reading>. The Select Normalized Reading screen is displayed.
the last reading taken is highlighted as the normalized reading. To select a different
normalized reading, touch to highlight the reading and then touch <OK>. The new reading
is selected and will serve as the normalized reading. The system returns to the Manual
Decay Test screen.
NOTE: You can also select the normalized reading while the Semi-Automated Decay Test is
underway. If you have not selected a normalized reading before selecting <End Test>, the
system automatically displays the Select Normalized Reading screen from which you can
select which reading to use as normalized.