224 ATOMLAB 500 DOSE CALIBRATOR
Also, the Wide Window Background CPM value is currently being exported (I didn’t realized this
before). It’s in the Wide Window record. Here is an example of the Wide Window record with a
brief description of the fields:
WideWindow, Cs-137, 40 DPM,14.0,1039 DPM,20000 DPM,Ok,35.24,25-800 keV
Where:
Cs-137: isotope name assigned for the wide window
40 DPM: wide window net count in DPM
14.0: wide window net count in CPM
1039 DPM: Background CPM for the wide window
20000 DPM: trigger level for the wide window
Ok: status of the wide window (Ok or Over)
35.24: total efficiency of the wide window
25-800 keV: ROI of the wide window
RESET
immediately re-starts the Atomlab firmware program.
SETID #-#
Changes a detector ID specified as “from-to”. Valid ID's are: 2, 4, 6, 8, 10, 12, and 14.
SMP {#}
Displays details on an inventory sample number. If no parameter is given, the number of stored
samples is displayed.
SSAV #
Sets the screen saver time in seconds. It's in effect only until the next reboot, then the
configured screen saver takes effect. If no parameter is given, it displays the current screen
saver time in effect.
SSRC {#}
Deletes the specified sealed source number. If no parameter is given, a summary of the current
defined sealed sources is displayed.
STAT {1/0}
This is for more advanced detector status viewing. This command will turn on an on-screen
display of detector status on the <Measure> (1=Set On, 0=Turn OFF). If no parameter is given, a
detailed state of the current connected detectors is displayed.
VER
Displays the Atomlab Display Unit's firmware versions as follows: UDM Vers: 1.22 10/12/2010
Windows CE 6.0 Based System