Biodex Medical Systems, Inc. © 2016 223
REQDATA
{Qatype} {Locationtype} {Date-Range}
Where:
QA Type: Set to “ALL” or 1 of 11 data types: CON, GEO, ACC, DEC, LIN, CAL, HVA,
CLB, BCK,
WIP, or CHI
Location Type: If the QA type is “Wip”, then location type can be “ALL” or AREA (Restricted
or
Unrestricted), SRC (Sealed Source), or PKG (Package)
Date-Range: “From-To” for a date range or a date for a single day. A date is in the format of
“mm/dd/yy” or can be “mm/dd” assuming the current year
Examples:
REQDATA W IP AREA 05/01-05/31
{get all the Restricted/Unrestricted wipe results for the month of May}
REQDATA BCK 05/22
{get all the wipe background results from May 22nd}
REQDATA ALL
{all recorded data}
REQDATA ALL 03/21/2009
{all data recorded on one day: 03/21/2009}
The QA Type’s are defined as:
CON - Daily and Expanded Constancy, Geo - Geometry, ACC - Accuracy,
DEC - Decay, Lin - Lineator,
CAL - Calicheck, HVA - wipe High Voltage Adjustment,
CLB - wipe Daily Calibration, BCK - wipe Background,
WIP - Wipe Tests, CHI - Chi-Square
The first item returned will be a CR/LF terminated line:
n Records
{count of records being sent}
Followed by the actual records in the current export format. If there are no records meeting the
criteria, all that is returned would be: 0 Records
For the Wipe Test Data, a line will be added immediately after the location line for the
Location Type:
Type,”Location Type” where ”Location Type” can be Restricted, Unrestricted, Sealed Source,
or
Package