92 ATOMLAB 500 DOSE CALIBRATOR
Figure 5.12. Two Reading Half Life Verification screen.
Figure 5.13. Half Life Verification Counting screen.
15. When the test is complete, the results are shown on the screen. The results include the
calculated Half Life and the variance percent from the published Half Life of the isotope.
Select <Print> to print your results. If you select <Back>, a message will be
displayed that if you proceed your data will be lost. Select <No> not to exit; select <Yes>
to exit and lose the test data.