P1 Patient Monitor User’s Manual
10-9
3. Set Wave Speed to the appropriate value. The larger the value, the faster
the scanning speed and the wider the waveform.
10.7.8. Setting Pitch Tone
The monitor can adjust the QRS tone according to the SpO
2
value. The pitch tone
function is on by default. The steps to turn off the pitch tone function are as follows:
1. Select the SpO
2
parameter area or waveform area to enter theSpO
2
menu.
2. SelectSpO
2
Setupsubmenu.
3. Switch offPitch Tone.
10.7.9. Setting PI Display
You can switch on or off PI display by following these steps:
1. Select the SpO
2
parameter area or waveform area to enter theSpO
2
menu.
2. SelectSpO
2
Setupsubmenu.
3. SetDisplay PIas toOnorOff.
10.7.10. Setting the Masimo SpO
2
When your monitor is equipped with Masimo SpO
2
module, you can set the following
contents:
10.7.10.1. Setting Sensitivity mode
To set the sensitivity mode, follow these procedures:
1.
Select the SpO
2
parameter area or waveform area to enter theSpO
2
menu.
2.
SelectSpO
2
Setupsubmenu.
3.
SelectSensitivityin the SpO
2
setting menu with the options ofMax,
NormalorAPOD.
Max: This mode should be used for the sickest patients, where obtaining
a reading is most difficult. The mode is recommended during procedures
and when clinician and patient contact is continuous.
Normal: This mode provides the best combination of sensitivity and
probe-off detection performance. The mode is recommended for the

Table of Contents