P1 Patient Monitor User’s Manual
10-10
majority of patients.
【APOD】: This mode is the least sensitive in picking up a reading on
patients with low perfusion but has the best detection for probe-off
conditions. The mode is useful for patients that are at particular risk of the
sensor becoming detached (pediatric, combative, etc.)
10.7.10.2. Setting Alarm Delay Time
To set the alarm time, follow these procedures:
1.
Select the SpO
2
parameter area or waveform area to enter the【SpO
2
】menu.
2.
Select【Alarm】submenu.
3.
Select【Alarm Delay】, and you can select the alarm delay time as required or
you can also set to【Off】.
10.7.10.3. Setting FastSat mode
The FastSat mode is clinically applicable during procedures when detecting rapid
changes in oxygen saturation is paramount such as induction, intubation and sleep studies.
To set the FastSat mode, follow these procedures:
1.
Select the SpO
2
parameter area or waveform area to enter the【SpO
2
】menu.
2.
Select【SpO
2
Setup】submenu.
3.
Select【On】or【Off】to enable or disable the 【FastSat】 mode.
10.7.10.4. Setting SmartTone
To set the SmartTone, follow these procedures:
1.
Select the SpO
2
parameter area or waveform area to enter the【SpO
2
】menu.
2.
Select【SpO
2
Setup】submenu.
3.
Select【On】or【Off】to enable or disable the 【SmartTone】.
When you set it to【On】, it will allow the audible pulse beep to beep when the
pleth shows signs of motion.
The pulse beep is suppressed during signs of motion when SmartTone is set to
【Off】.
10.7.10.5. Setting SpHb mode
While monitoring Hemoglobin levels, there are two blood sample sources from which