P1 Patient Monitor User’s Manual
15-7
15.2.7. Events Review Page
The monitor stores alarm events and system events in real time. Alarm event types
include physiological alarm event. When an alarm event occurs, the monitor will store the
values of relevant parameters at the time of occurrence and the relevant waveforms for 16
seconds before and after the time of occurrence.
CAUTION:
A sudden loss of power has no impact on the events stored.
15.2.7.1. Entering the Events Review Page
Choose one of the following methods to enter the events review page:
Select【Review】quick keySelect【Event】submenu.
Select【Main Menu】quick keyfrom【Review】column to select【Events】.
The event review page displays a list of events in the order in which they occurred.
The most recent event is displayed at the top. The number of asterisk symbols (*) before
an event an event matches alarm priority.
The event identifier on the left side of the alarm event displays different types of
events with different color blocks:
Red: high priority alarm event
Yellow: medium priority alarm event
Cyan: low priority alarm event
The number of currently events and the total number of filtered events are displayed
at the top right corner of the event list. For example, 3/10 indicates that there are a total of
10 selected events, and currently there are 3 events.
15.2.7.2. Configuring the Filter
You can filter events by time, alarm priority, parameter category or event type. To
configure the filter, follow this procedure:
1. Enter events review page to switch on 【Filter】.
2. Select 【Filter Setup】 and set the desired filter criterion. Events after
filtering will be displayed in the event list.