P1 Patient Monitor User’s Manual
15-8
CAUTION:
If Filter is not switch on, the relevant setting in Filter Setup will not take
effect.
15.2.7.3. Viewing Event Details
To view waveforms and parameter values at the selected event time, follow this
procedure:
1. Enter the event review page.
2. Select Detail. You can perform the following operations on this page:
Select to enter setup menu. To set Wave Speed, Record
and ECG Gain.
Select Overview to return to the compressed waveform page.
CAUTION:
Please ensure that the best ECG lead with largest waveform amplitude and the
highest signal-to-noise ration is selected. Choosing the best ECG lead is very
important to recognize cardiac beat, classify cardiac beat and recognize
ventricular fibrillation.
15.2.8. Full Disclosure Review Page
On the Full Disclosure review page, you can review waveform data 72 hours. You
can view compressed waveforms, full waveforms and numeric values.
15.2.8.1. Entering the Full Disclosure Review Page
Choose one of the following methods to enter the Full Disclosure review page:
SelectReviewquick keySelectFull Disclosuresubmenu.
SelectMain Menuquick keyform Review column select Full
Disclosure.