P1 Patient Monitor User’s Manual
16-8
Input Parameter Unit Full Name
CaO
2
mL/dL, mL/L arterial oxygen content
CvO
2
mL/dL, mL/L
venous oxygen content
VO
2
mL/min
oxygen consumption
RQ Respiratory quotient
ATMP mmHg, kPa atmospheric pressure
Height cm, inch height
Weight kg, lb weight
16.5.3. Output Parameters and calculation formula
Output
Parameters
Unit Full Name Formula
BSA m
2
body surface area
BSA=HT
0.725
×WT
0.425
×
0.007184
C(a-v)O
2
mL/L,
mL/dL
arteriovenous
oxygen content
difference
C(a-v)O
2
=CaO
2
×CvO
2
O
2
ER %
oxygen extraction
ratio
O
2
ER=CaO
2
CvO
2
/ CaO
2
DO
2
mL/min oxygen transport DO
2
=C.O.×CaO
2
PAO
2
mmHg,
kPa
partial pressure of
oxygen in the
alveolar
PAO
2
=FiO
2
×(ATMPwater
pressure) 】-(PaCO
2
×1.25)
Wherein the water pressure is
selected to be 47mmHg (6.3kPa)