P1 Patient Monitor User’s Manual
13
No. Full Name Prompt information
18 VentricularBigeminy Bigeminy
19 VentricularTrigeminy Trigeminy
20 Frequent PVCs Frequent PVCs
21 Premature ventricular contraction PVC
22 Missed Beat Missed Beat
23 Atrial Fibrillation A-Fib
24 Atrial Fibrillation End A-Fib End
25 ECGNoise ECG Noise
26 Irregular Rhythm Irregular Rhythm
27 Irregular Rhythm End Irregular RhythmEnd
A.8.2 RESP Specifications
A.8.2.1 Measurement specification
Measurement
parameter
Respiration Rate and respiration waveform
Source
RA-LA, RA-LL (default)
Excitation waveform
Excitation frequency: 64 kHz; Error: ±10%
Excitation current
0.3mA RMS
Respiration Apnea
Alarm
Fixed high priority alarm
Adjustable delay time: 10~60s, error ±3s or ±10%, whichever
is greater.
Cardiac interference
alert
Fixed high priority alarm
RR measurement
range and accuracy
Measurement range 0~150 rpm
Resolution 1 rpm
Accuracy
±2 rpm or ±2% of reading, whichever
is greater.
Bandwidth
0.2 Hz ~2.5Hz (-3dB~+0.4dB)
Sweep speed
6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s, error is within ±10%.