P1 Patient Monitor User’s Manual
14
Baseline impedance
range
200~2500 (using defibrillator proof cable with resistance of
1k)
Measuring impedance
range
0.3~3
Gain
×0.25, ×0.5, ×1, ×2, ×4
A.8.3 NIBP Specifications
A.8.3.1 Standard
Meet standard of IEC80601-2-30
A.8.3.2 Measurement Specification
BLT NIBP
Measurement
parameters
SYS, DIA, MAP, PR
Mode of operation
Manual, Auto, STAT, Sequence
Periodic measurement
The monitor automatically starts NIBP measurement at set
intervals.
Sequence mode
At least 5 groups are supported, and each group individually
sets the interval and number of periodic measurement.
STAT mode cycle
time
5 min
Measurement range
of cuff pressure
0~300 mmHg
Initial inflation
pressure
Adult: 100~280mmHg, default 160mmHg
Pediatric: 100~240 mmHg, default 130mmHg
Neonatal: 60~140mmHg, default 100mmHg
Sensor calibration
time
One year (recommend)
Unit
mmHg, kPa
Dynamic pressure
measurement range
Systolic
Adult: 30~270 mmHg (4.036.0 kPa)
Pediatric: 30~235 mmHg (4.031.3 kPa)