P1 Patient Monitor User’s Manual
15
Neonatal: 30~135 mmHg (4.0~18.0 kPa)
Diastolic
Adult: 10~220 mmHg (1.3~29.3 kPa)
Pediatric: 10~220 mmHg (1.3~29.3 kPa)
Neonatal: 10~110 mmHg (1.3~14.6 kPa)
Mean
Adult: 20~235 mmHg (2.7~31.3 kPa)
Pediatric: 20~225 mmHg (2.7~30.0 kPa)
Neonatal: 20~125 mmHg (2.7~16.6 kPa)
Dynamic Pressure
Measurement Error
of Simulator
±8 mmHg (±1.1kPa)
Static pressure
accuracy
±3 mmHg (±0.4kPa)
Pressure resolution
1 mmHg or 0.1kPa
PR measurement
range and accuracy
Measurement
range
40 ~ 240 bpm
Accuracy ±3bpm or ±3%, whichever is greater
Maximum
measurement time
Neonatal: <90s
Adult, Pediatric: <120s
First overvoltage
protection point
Adult: 297±3 mmHg
Pediatric: 252±3 mmHg
Neonatal: 147±3 mmHg
Second overvoltage
protection point
Adult: 315±10 mmHg
Pediatric: 265±10 mmHg
Neonatal: 155±10 mmHg
Note: The accuracy of NIBP cannot be determined by using a simulator, but under many
conditions, it is still necessary to use a simulator to test its performance (for example, a
simulator is required for quality control in the production process), and the simulator of
the model specified by the manufacturer shall be used for this performance test.
Clinical index
Evaluation method for
Follow ISO 81060-2 standard, in which