P1 Patient Monitor User’s Manual
16
clinical accuracy of blood
pressure
Adult (including pediatric): auscultation method
Neonatal: invasive method
Accuracy
Systolic and diastolic pressures: mean error: ±5mmHg,
standard deviation: 8 mmHg
Mean pressure: not participating in evaluation
Overall measurement time
20~45s (typical value)
Suntech NIBP (optional)
Way of
measurement
Oscillometric.
Range of
measurement
Systolic
Adult: 40~260 mmHg (5.3~34.7 kPa)
Pediatric: 40~160 mmHg (5.3~21.3 kPa)
Neonatal: 40~130 mmHg (5.3~17.3 kPa)
Diastolic
Adult: 20~200 mmHg (2.7~26.7 kPa)
Pediatric: 20~120 mmHg (2.7~16.0kPa)
Neonatal: 20~100 mmHg (2.7~13.3kPa)
Mean
Adult: 26~220 mmHg (3.5~29.3 kPa)
Pediatric: 26~133 mmHg (3.5~17.7 kPa)
Neonatal: 26~110 mmHg (3.5~14.7kPa)
Static pressure
accuracy
±3 mmHg (±0.4kPa)
Unit
mmHg, kPa
Pulse rate range
30 ~ 220 bpm
Pulse Rate Accuracy
±2bpm or ±3%, whichever is greater
Inflation time for
cuff
<75s
Measurement
protection time
Adult: <180s
Pediatric: <180s
Neonate: <90s