P1 Patient Monitor User’s Manual
17
Initial inflation
pressure
Adult: 120~280mmHgdefault 160mmHg
Pediatric: 80~170mmHgdefault 120mmHg
Neonate: 60~140mmHgdefault 90mmHg
Overpressure
Protection
Adult
Pediatric
Neonate
Hardware and software double protections
<300 mmHg
<300 mmHg
<150 mmHg
Clinical accuracy
Systolic and diastolic pressures: mean error: ±5mmHg, standard
deviation: 8 mmHg
A.8.4 SpO
2
Specifications
A.8.4.1 Standard
Meet the standard of ISO 80601-2-61
A.8.4.2 Specification
BLT SpO
2
Measurement parameter
Measure two channels of SpO
2
, PR, PI,
SpO
2
and RR,
and show SpO
2
waveform and respiration waveform.
Sensitivity
High, Medium, Low
Measurement range and
accuracy
Measurement range 0~100%
Clinical accuracy
70%~100%: 3% (SpO
2
probe
included in Appendix);
0~69%: unspecified
Simulator
measurement error*
70%~100%: ±2% (non-motion
conditions)
0~69% unspecified
SpO
2
Update period
Normal 2s
Maximum 25s
SpO
2
Response time
Sensitivity High 8s