P1 Patient Monitor User’s Manual
20
Accuracy
±0.1% or ±10% of reading,
whichever is greater
Update period
Normal 2s
maximum 25s
Masimo SpO
2
Measurement
parameter
SpO
2
, SpMet, PVI, SpHb, SpOC, PI and SpCO
Measurement range
0% to 100%
Resolution
1%
Accuracy
70% to 100%:±2%adult/pediatric, non-motion conditions
70% to 100%:±3%neonate, non-motion conditions
0% to 69%,unspecified
Average time
2-4s, 4-6s, 8s, 10s, 12s, 14s, 16s
Alarm range and
error
Alarm range
0% to 100%, high/low limit can be
adjusted continuously.
Alarm error ±1%
PR
Measurement range 25 bpm to 240 bpm
Accuracy ±3 bpm (non-motion conditions)
Alarm range and
error
Alarm
range
0bpm to 300bpm, high/low
limit can be adjusted
continuously.
Alarm
error
±1bpm
SpCO
Measurement range 0% to 100%
Accuracy
0% to 40%±3%(non-motion conditions)