P1 Patient Monitor User’s Manual
21
>40%, unspecified
Resolution 1%
SpMet
Measurement range 0.0% to 100.0%
Accuracy
0% to 15%:±1%(non-motion conditions)
>15%unspecified
Resolution 0.1%
PI
Measurement range At least 0.05% to 20.00%
Resolution 0.01%
Accuracy
±0.1% or ±10% of reading, whichever
is greater
SpHb
Measurement range 0.0 g/dl to 25.0 g/dl
Accuracy
8 g/dl to 17 g/dl: ±1 g/dl (non-motion
conditions)
>8 g/dl or >17 g/dl, unspecified
Resolution 0.1g/dl
A.8.5 TEMP Specifications
A.8.5.1 Standard
Meet the standard of ISO 80601-2-56.
A.8.5.2 Measurement Specification
Parameter
T1,T2,T
D
Probe
YSI400 series probe (2252@25, accuracy ±0.1ºC)
Measurement site
Surface and coelom
Measurement range and
accuracy
Measurement
range
0.0ºC ~50.0ºC (32.0℉~122.0)