P1 Patient Monitor User’s Manual
22
Resolution
0.1ºC or 0.1℉
Accuracy of
circuit
±0.1℃ (±0.2℉) (without sensor)
Alarm range
0.0℃ to 50.0℃ (32.0℉~122.0℉), high/low limit can
be adjusted continuously.
Alarm error
±0.1℃ or ±0.1℉
Updated time
Every about 1~2s
Minimum measurement
time
Surface: 100s
Coelom: 80s
A.8.6 IBP Specifications
A.8.6.1 Standard
Meet the standard of IEC 60601-2-34.
A.8.6.2 Functional specification
Measurement
parameters
Dual channels IBP parameters (including systolic blood
pressure, diastolic blood pressure, average pressure, PR) and
waveforms
Scale
Manual, interval and automatic scale settings
Unit
mmHg, kPa, cmH
2
O
PPV
Measurement
range
0%~50%
Resolution 1%
Static pressure
measurement range
and accuracy
Measurement
range
-6.7kPa ~+48.0kPa (-50mmHg ~+
360mmHg)
Resolution 1 mmHg
Accuracy
±0.3kPa (±2mmHg) or ±2%, whichever is
greater (without sensor)
Dynamic pressure
measurement range
Measurement
range
-6.7kPa ~+48.0kPa (-50mmHg ~+
360mmHg)