P1 Patient Monitor User’s Manual
23
and accuracy
Accuracy
±0.3kPa (±2mmHg) or ±2%, whichever is
greater (without sensor)
Frequency response
Including sensor
0Hz10Hz
Only host
0Hz12Hz
IBP zero range
-200mmHg+200mmHg
PR
Measurement
range
30bpm ~300bpm
Resolution 1bpm
Accuracy ±1% or ±1bpm, whichever is greater
Pressure sensor
Nominal
sensitivity
5uV/V/ mmHg
Output
impedance
300~3000
Volumetric
displacement
<0.04 mm
3
/100 mmHg
Error ±2%
IBP analog output
delay
35ms
A.8.7 CO
2
Specifications
A.8.7.1 Standard
Meet the standard of ISO 80601-2-55.
A.8.7.2 Functional Specification
Measurement
parameter
EtCO
2
, FiCO
2
, a CO
2
waveform and awRR
Measurement
method
Mainstream, Sidestream/Microflow
Unit
mmHg, kPa and %