A.8.7.3 Performance Specification
EtCO
2
/FiCO
2
measurement range
0%~19.7% (0mmHg~150mmHg)
EtCO
2
/FiCO
2
measurement accuracy
± (0.43% + 8% of reading)
Accuracy drift
Meets the requirements for measurement accuracy within 6
hours.
EtCO
2
/FiCO
2
display
resolution
0.1% or 1mmHg
awRR measurement
range
0~150 bpm
awRR measurement
accuracy
±1 bpm
awRR display
resolution
1 bpm
awRR alarm range
0bpm to 150bpm, high/low limit can be adjusted continuously.
Sampling frequency
and accuracy of gas
(only sidestream)
C1
50 mL/min~200mL/min, ±10%, can
be adjusted.
Other sidestream CO
2
50 mL/min±10mL/min
Response time
Sidestream CO
2
<3s
Mainstream CO
2
<1s
A.8.7.4 The effects on CO
2
measuring values caused by the interfering gases
BLT TL _ Si & BLT CPT
The accuracy of CO
2
is affected by interfering gases and water vapor. For example,
N2O, a halide-containing anesthetic gas can raise the CO
2
reading (about 2%-10%), and
helium and oxygen can reduce the CO
2
reading (1%-10%), so in the presence of
interfering gas, the user should send relevant command to the module (the instrument’s
compensation menu to adjust the interference gas data), so that the module (instrument)
can meet the nominal accuracy requirements.