P1 Patient Monitor User’s Manual
27
Alarm limit Range Step
NIBP-M-High Adult: (low limit+1mmHg) ~235 mmHg
1mmHg
Pediatric: (low limit+1 mmHg) ~225 mmHg
Neonatal: (low limit+1 mmHg) ~125 mmHg
NIBP-M-Low 20 mmHg ~ (high limit-1 mmHg)
NIBP-D-High Adult: (low limit+1mmHg) ~220 mmHg
1mmHg
Pediatric: (low limit+1 mmHg) ~220 mmHg
Neonatal: (low limit+1 mmHg) ~110 mmHg
NIBP-D-Low 10 mmHg ~ (high limit-1 mmHg)
SunTech NIBP
Alarm limit Range Step
NIBP-S-High
Adult: (low limit+1 mmHg) ~260 mmHg
1mmHg
Pediatric: (low limit+1 mmHg)~160 mmHg
Neonatal: (low limit+1 mmHg) ~130mmHg
NIBP-S-Low 40 mmHg ~ (high limit-1 mmHg)
NIBP-M-High Adult: (low limit+1mmHg) ~220 mmHg
1mmHg
Pediatric: (low limit+1 mmHg) ~133 mmHg
Neonatal: (low limit+1 mmHg) ~110 mmHg
NIBP-M-Low 26 mmHg ~ (high limit-1 mmHg)
NIBP-D-High Adult: (low limit+1mmHg) ~200 mmHg
1mmHg
Pediatric: (low limit+1 mmHg) ~120 mmHg
Neonatal: (low limit+1 mmHg) ~100 mmHg
NIBP-D-Low 20 mmHg ~ (high limit-1 mmHg)
A.9.4 SpO
2
Alarm Specification
Alarm limit Range Step
SpO
2
High
(low limit+1%) ~100%
1%
SpO
2
Low
(SpO
2
Desat +1%) ~ (high limit-1%)
SpO
2
Desat
0%~ (low limit-1%)
PR High (PR low limit+1bpm)~350bpm
1bpm