P1 Patient Monitor User’s Manual
28
PR Low 0bpm~ (PR high limit-1bpm)
△SpO
2
High
(low limit+1%) ~100%
1%
△SpO
2
Low
0bpm~ (high limit-1%)
RR High
(low limit+1%) ~150bpm
1bpm
RR Low
0bpm~ (high limit-1bpm)
A9.5 TEMP Alarm Specification
Alarm limit Range Step
T1/T2 High (low limit+0.1ºC) ~50.0ºC 0.1 ºC
T1/T2 Low 0 ºC~ (high limit-0.1ºC) 0.1 ºC
TD High 0ºC~5.0ºC 0.1 ºC
A.9.6 IBP Alarm Specification
Alarm limit Range Step
IBP-M-High (low limit +1mmHg) ~360mmHg
1mmHg
IBP-M-Low -50mmHg~ (high limit -1mmHg)
IBP-D-High (low limit +1mmHg) ~360mmHg
1mmHg
IBP-D-Low -50mmHg~ (high limit -1mmHg)
NIBP-S-High (low limit +1mmHg) ~360mmHg
1mmHg
NIBP-S-Low -50mmHg~ (high limit -1mmHg)
A.9.7 CO
2
Alarm Specification
Alarm limit Range Step
Apnea delay time 20 s~60 s 5s
EtCO
2
High (low limit+1mmHg) ~152mmHg
1 mmHg
Et CO
2
Low 0mmHg~ (high limit-1mmHg)
Fi CO
2
High 0~152mmHg 1 mmHg
awRR High (low limit+1bpm) ~150 bpm
1
bpm
awRR Low
0bpm~ (high limit-1bpm)