P1 Patient Monitor User’s Manual
3-14
light of all other alarms.
The monitor suppresses all system prompt tone, including heart beat tone,
pulse tone, etc.
WARNING:
In Privacy mode, all audible alarms are suppressed and the alarm light is
deactivated at the monitor. Alarms are presented only at the
Central
monitoring system. Pay attention to potential risk.
CAUTION:
Cannot enter the privacy mode if a low battery alarm occurs.
3.6.6.2. Exit the Privacy Mode
The monitor automatically exits the privacy mode in any of the following situations:
The monitor disconnects from the Central Monitoring System.
The low battery alarm occurs.
You can also press 【Exit Privacy Mode】on the screen to manually exit the privacy
mode.
3.6.7. Outdoor Mode
The outdoor mode is intended for transferring patients outdoors. The monitor behaves
as follows after entering the outdoor mode:
The parameter color is white and unchangeable.
The screen brightness is automatically changed to 10.
3.6.7.1. Entering the Outdoor Mode
The methods of entering the outdoor mode are as below:
1. Select 【Main Menu】 quick key.
2. Select 【Outdoor Mode】 on the 【Display】 column.