P1 Patient Monitor User’s Manual
7-7
7.7.3.3. Setting Alarm Tone Mode
The setting steps are as follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuSoundsubmenuAlarm Sound, you can
selectISO.
7.7.3.4. Setting Alarm Tone Interval
You can set the alarm tone interval. The steps are as follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuSoundsubmenu.
3. Set High Alarm Interval, Med Alarm Interval and Low Alarm
Interval.
High Alarm Interval
:
3
15s, and the default value is 10s.
Medium Alarm Interval
: 3
30s,
and the default value is 20s.
Low Alarm Interval
: 16
30s
and the default value is 20s.
7.7.3.5. Setting Reminder Volume
When the alarm volume is 0, alarm reset or the alarm is off, the monitor can provide
periodic alarm prompt tone to remind you that there is still an activated alarm in the
current system. This function is turned on by default.
You can set the alarm tone as follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmbuttonPause /Resetsubmenu.
3. Set Alarm Pause duration. You can set Alarm Pause durationto
1min, 2min, 3minorPermanent, the default is2min.
4. SetAlarm Off Reminderswitch.
On: The monitor provides an alarm tone according to the set interval.