P1 Patient Monitor User’s Manual
7-8
Off: The monitor does not provide an alarm tone.
5. Set Reminder Interval. You can set Reminder Intervalto 1min,
2min, 3min, 5minor10min, the default is5min.
7.7.3.6. Setting Alarm Tone Enhancement
The monitor provides an alarm tone enhancement function. If the alarm exceeds the
set time and is not confirmed, the alarm volume can be automatically enhanced.
The steps to set the alarm tone enhancement are as follows
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmSound.
3. SetAuto Increase Volumeto3 Steps, 2 Steps, 1 Stepsor
Off.
3 Steps:
After the alarm occurs, the alarm volume will be
automatically increased to level 3, if the set time is not confirmed.
2 Steps : After the alarm occurs, the alarm volume will be
automatically increased to level 2, if the set time is not confirmed.
1 Steps : After the alarm occurs, the alarm volume will be
automatically increased to level 1, if the set time is not confirmed.
Off: After the alarm occurs, the set time is not confirmed, and the
alarm volume remains unchanged.
4. Set
Increase Volume Delay
, select sound enhanced delay time.
7.7.4. Setting Alarm Delay Time
For the over-limit alarm of continuous measurement parameters, the alarm delay
time can be set. If the condition of triggering alarm disappears within the delay time, the
monitor will not alarm.
Set the delay time for the alarm by following these steps:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuOthersubmenu.