19
8.1 Fuse replacement
Ź Use a small screwdriver to release the clip on the fuse holder.
Ź Pull out the fuse holder.
Ź Replace both fuses.
Ź Reinsert the fuse holder in such a way that it audibly snaps into place.
Note
If the fuses burn out frequently, the device is faulty and should be
repaired, see Technical Service.
9. Technical Service
Ź For service and repairs, please contact your national B. Braun/Aesculap
agency.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone: +49 (7461) 95 -1601
Fax: +49 (7461) 14 -939
E-Mail: ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone: +1 (800) 214 -3392
Fax: +1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
Unrealistic color representation on
the monitor
White balance failed Start white balance-adjustment
Monitor color setting misadjusted Reset to factory setting
Video cable defective Replace video cable
The LED light source is no longer
controlled via the camera head
Problems with the communication interface between
the light source and camera control unit
Connect the communication interface cable correctly
Remove the communication interface cable and man-
ually operate the light source
“System error” appears on the
monitor OSD
Internal product error Restart camera
“System error” appears on the
monitor OSD even after restart
Internal product error Send system to service, see Technical Service.
Malfunction Cause Remedy
DANGER
Risk of fatal injury from electric shock!
Ź Unplug the device before changing the fuses!
WARNING
Risk of injury and/or malfunction!
Ź Do not modify the product.
CAUTION
Malfunction of the product due to damage in
transport!
Ź Select suitable and safe packaging (ideally the
original packaging).