144 (167) BRUKER Installation Manual Version 001
Sample Heater Option
Figure 13.12.B-ACS Heater Power Supply Wiring PK240
A
G
B
F
C
E
D
H
GND
L
N
Power Line Filter
& Main Fuse
2x1.6A/250V
time lag
PE
1
2
2
1
2a
1a
I
24
32
14
II
16
20
28
18
22
26
30
PE Rear Panel
PE Front Panel
PE Chassis
Switch
Power Supply
Connector Pk240
DC Connector
on Rear Panel
B-ACS Heater Power Supply H1455 from Ser. No. 35
1a
2a
2
1
bl
brn
230V AC N
230V AC L
0.75 qmm
0.75 qmm
Line Filter Switch
28
30
1
2
bl
brn
230V AC N
230V AC L
0.75 qmm
0.75 qmm
Switch
Connector
Power Supply
C
A
12
10
yel
rd/brn
GND
+ 26V DC
1.5 qmm
1.5 qmm
E
G
16
14
yel
rd/brn
- Sense
+ Sense
0.75 qmm
0.75 qmm
Connector
Power Supply
DC Connector
on Rear Panel
PE
PE
yel/grn
yel/grn
PE
PE
1.5 qmm
1.5 qmm
PE Chassis
Line Filter
32
PE
yel/grn
yel/grn
PE
PE
1.5 qmm
1.5 qmm
Connector
Power Supply
Front Panel
Rear Panel

Table of Contents