Communication Errors
Installation Manual Version 001 BRUKER 53 (167)
12. Sample Detect at Carrousel Failed
The optical detector at the magazine sees no sample.
13. Sample Detect at Magnet Failed
The optical detector at the magnet sees no sample within 30 seconds of lift
on.
14. Sample Grasping Failed
The sample could not be removed.
16. CDW Detection Failed
The number read from the internal code wheel is not within the allowed
range between 1 and 60.
21. CDW MISADJUSTED FOR POSITION No: XX
Code XX could not be read from the code wheel.
22. No Free Magazine Position
No free magazine position could be found for the sample in the pincher.
23. Sample Missing
No sample could be found in the present magazine position.
25. Failure of Spinning Device
The mechanism of the bar code reader is no in its required idle position.
26. Arm Positioning Failed
The exact outer magazine position (61 to 120) could not be reached.
If one of the above mentioned failures is observed the program branches off into
an error correction routine. It can be restarted by pressing the CONT button or by
giving the command „CO“ via the RS 232C link. As an alternative the command
„HO’’ can be used to move the sample changer into its „HOME’’ position.
Communication Errors 7.3
Failures in the communication between the Sample Changer and the Host Com-
puter can occur if an incorrect command is given by the host computer, or if un-
connected hardware is addressed. Following such an error message, the sample
changer will accept a new command and not enter the error mode. Error messag-
es are not displayed but are sent back to the host computer. If any of these errors
occur consult your operating instructions, or a BRUKER service representative.

Table of Contents