28 (167) BRUKER Installation Manual Version 001
Settings and Adjustments
Figure 5.1. Cabinet View Showing Boards and Pneumatic Assembly
Make sure the mains connector is plugged into the connection panel of the square
column piece (see Figure 4.12.
). Pull the red security switch outwards and push
the black start switch (Figure 5.2.
). All mechanical functions can now be switched
manually by pressing the push buttons on the pneumatic assembly (Figure 5.1.
).
Figure 5.2. Security Switch
1 VERTICAL SENSORS
2 HORIZONTAL SENSORS
3 HORIZONTAL
POSITION
4 MAGAZINE POSITION
5 LIGHT BARRIER (at the)
SHIM SYSTEM
6 LIGHT BARRIER
(at the) MAGAZINE
CONNECTION
TO LCD DISPLAY
CONNECTION
TO OPTICAL
CODE WHEEL
POWER SUPPLY
CPU
INTERFACE
RS232
RESERVE (Bar Code Option)
TERMINATOR
= SAMPLE LIFT VALVE
= MOTION DIRECTION
= MAGAZINE MOTION
= ARM TO MAGAZINE
= PINCHER OPEN
= ARM DOWN
= ARM UP
= ARM TO MAGNET
= RESERVE (Bar Code Reader Option)