18 (167) BRUKER Installation Manual Version 001
Mounting Instructions
Figure 4.5. Column Assembly
Mounting the Column Assembly to the Magnet 4.7
Fasten the two top adjustable arm assemblies to the round portion of the column.
For Magnets without N2 Towers:
Remove the bolts and brackets (see Figure 4.6.
) from the magnet end of the ad-
justable arms and place them next to the magnet lifting lugs.
Lift the column assembly upwards so that the cabinet is on the top. Secure the
two top adjustable arms to the lifting lugs of the magnet with the bolts and brack-
ets.
CABLES AND PNEUMATIC
HOSE (From Arm to Cabinet)
CABLES AND PNEUMATIC
HOSE (From Connection Panel to
Cabinet)
CLAMP COVER
SQUARE CLAMP
VERTICAL LOCKING
MECHANISMS
CABLE AND PNEUMATIC
HOSE CONNECTION PANEL
COLUMN PIECES
Screws
SIDE OPENING