Installation Manual Version 001 BRUKER 87 (167)
The M120 optic is connected to the lower connection (4 Pol.) on the back of the
light barrier cabinet (figure 11.17
).
Adjustment Procedures
1. Adjustment of the angle of inclination (tilt):
The magazine belt position should be unoccupied.
The Distance Adjustment Screw (figure 11.18
) is set on maximum „FAR“.
With a reflective background a reduction of the distance adjustment is pos-
sible. The Distance Adjustment Screw is turned counter clockwise to
„NEAR“ until the display „OPE“ is no longer activated (this provide an in-
crease in the aperture of the background).
The optic is first adjusted as far as possible to the bottom, until the maga-
zine belt is detected, then turned back gradually until the belt can no longer
be detected.
You must ensure that the M120 optic detects a sample only in the outside
magazine belt position, not the inside position.
2. Adjustment of the distance:
The magazine belt position should be occupied.
The Distance Adjustment Screw is set on minimum „NEAR“.
Turn the Distance Adjustment Screw clockwise to „FAR“ until the display is
activated.
Note:
Excessive dust or dirt at the lens surface of the optics reduces the optics
recognition performance. Keep these surfaces clean.
Light Barrier at Shim System 11.3.5
Figure 11.23.Light Barrier Magazine Board H1288
33µ10VSAL
470
10K
10K
470
10µ25VSAL
100nF50V
DC/DC
H1288
LB60
LB120
V5R15
LED1
LED2
1
3
4
2
1
1 Red
2 Black
3 Green
4 White
1 Brown
2 Blue
3 Pink
4 Black
5 Free
Front View
5
4
2
3
1
Front View
1 Brown
2 Yellow
3 White
4 Green
1 2 3 4
+
+