THS – FI022GB2K8v3 --- V - MAINTENANCE
60
If the test fails, one of the following messages will be seen at point 4:
Display Comment
Test FE NULL <
Test SS
Test Non-FE
The sample is considered too large to carry out a valid test,
with a signal amplitude >400% of the SE alarm threshold.
The result of the test is recorded as NO.
Test FE NO <
Test SS
Test Non-FE
Possible causes:
- No transit has been detected for 20 seconds
- The signal from the sample is below the SE alarm
threshold
- Three consecutives passages were giving negative result
- The system of ejection is enabled, but ejection has not
been confirmed via the “ejection confirmation” input
The outcome of the test, whether positive or negative, is always stored in memory.
In the case of abnormal operation, carry out a test using the sample indicated on the Configuration Sheet enclosed with each unit after selecting the
“default” product.