THS – FI022GB2K8v3 --- V - MAINTENANCE
73
Change Password menu
Remote Possible settings Standard setting Parameter
Operator
US
6 alphanum. characters 000001
Global
Description Models
Operator’s password
Select each character via the (increase) e (decrease) keys and confirm by pressing the key.
All models
Remote Possible settings Standard setting Parameter
Supervisor
SU
6 alphanum. characters 000002
Global
Description Models
Supervisor’s Password
Select each character via the (increase) e (decrease) keys and confirm by pressing the key.
All models
Remote Possible settings Standard setting Parameter
Engineer
EG
6 alphanum. characters 000003
Global
Description Models
Engineer’s Password
Select each character via the (increase) e (decrease) keys and confirm by pressing the key.
All models
Remote Possible settings Standard setting Parameter
Q-C Op.
QU
6 alphanum. characters 000004
Global
Description Models
Quality control operator’s password
Select each character via the (increase) e (decrease) keys and confirm by pressing the key.
All models
Remote Possible settings Standard setting Parameter
Q-C Head
QH
6 alphanum. characters 000005
Global
Description Models
Head of quality control’s password
Select each character via the (increase) e (decrease) keys and confirm by pressing the key.
All models
Remote Possible settings Standard setting Parameter
Remote Prog.
RM
6 alphanum. characters 000006
Global
Description Models
Password for remote programming
Select each character via the (increase) e (decrease) keys and confirm by pressing the key.
All models
ALM Configuration menu
Remote Possible settings Standard setting Parameter
Input logic
LO
P / N P
Global
Description Models
Input logic
Applies to all inputs. If set to P (positive logic) all inputs are read as active when they are pulled to +V (with pull-down
incorporated on card ALM).
If set to N (negative logic) all inputs are read as active when they are pulled to -V (with pull-up incorporated on card
LM).
All models
Remote Possible settings Standard setting Parameter
BIN_FULL input
IL
NC / NO NO
Global
Description Models
Type of device connected to the “BIN_FULL” input
NC: Normally Closed; NO: Normally Open
If set to NO the Metal Detector gives an alarm when the BIN_FULL input is closed.
If set to NC the alarm is given when the input is open.
All models