THS – FI022GB2K8v3 --- V - MAINTENANCE
72
Remote Possible settings Standard setting Parameter
Transmission
TX
ON / OFF ON
Global
Description Models
Enabling of the transmitter
If set to OFF it is possible to understand if a noise is electrical or mechanical. Refers to the Instructions manual.
When TX=OFF the THS does not detect metals. The parameter is automatically set to ON if the THS is switched off
and on again.
All models,
except THS/MN
Remote Possible settings Standard setting Parameter
TX channel
CH
1 / 2 1
Global
Description Models
Transmission channel
Selection of different channels on adjacent detectors for automatic synchronisation. This parameter is not available
on some models.
All models,
except THS/MN
Remote Possible settings Standard setting Parameter
Display contrast
DC
0 – 50 0
Global
Description Models
Display contrast
Display contrast can vary with changes in environmental temperature.
All models
Remote Possible settings Standard setting Parameter
Time
TM
HH:MM -
Global
Description Models
Current time
HH = Hours; MM = Minutes. In remote programming #TM=HH:MM
All models
Date settings menu
Remote Possible settings Standard setting Parameter
Year
-
Last two figures
of the year
-
Global
Description Models
Year
For remote programming, see parameter DA.
All models
Remote Possible settings Standard setting Parameter
Month
-
1 - 12 -
Global
Description Models
Month
For remote programming, see parameter DA.
All models
Remote Possible settings Standard setting Parameter
Day
-
1 - 31 -
Global
Description Models
Day
For remote programming, see parameter DA.
All models